The Sable Systems FC-10 is
reliable, research-grade, and field-portable. The FC-10
boasts resolution and stability superior to any other
single channel fuel cell based oxygen analyzer.
Advantages. Sable Systems’ FC-10 equals or improves upon
the noise, drift and resolution performance of heated-zirconia
analyzers without their troublesome high power
consumption, fragility, systematic errors, and
exorbitant cost. The FC-10 is one of Sable Systems’ line
of Field-Aware instruments; it is compact and light,
powered by uncritical 8 - 20VDC power, requires only a
minute of warm-up time, and consumes less than two watts
of power

FC-10
Specifications
• Range: 1% to 100% oxygen, <1 to 105+ kPa oxygen
• Flow Rates: 10 - 500 ml/min recommended
• Panel Display: Digital 2-line alpha numeric LCD,
simultaneous display of oxygen (0.0001 % or 0.0001 kPa
resolution) and barometric pressure (0.0001 kPa
resolution)
• Voltage Outputs: 4, with multiple ranges on each
• Serial Output: Percent oxygen, kPa oxygen, both to
(0.0001%/kPa), barometric pressure (to 0.0001 kPa) and
fuel cell temperature (to 0.001°C) with comma
delimiters; resolution 1 part in 1,000,000 for
concentration and pressure.
• Power Requirement: 8 - 20 VDC at approx. 2 W, reverse
polarity protected. UL/CSE 120 VAC to 12VDC. AC adapter
is supplied. (220VAC to 12VDC adapter available upon
request)
• Accuracy and Resolution: At constant temperature,
accuracy better than 0.1%, 2 - 100% oxygen;
Resolution 0.0001%; Please note the difference between
accuracy and resolution
• Noise and Drift: At constant temperature, flow rate
and oxygen concentration: Noise < 0.002% pk-pk over 20
minutes; Drift < 0.01% over 24 hours
• Temperature Compensation: Internal to the fuel cell,
optimized over 15 – 45 °C range; External
optionally uses temperature data (full details supplied
on request)
• Size/Weight: 13" x 10" x 4" (33 x 25 x 10 cm); less
than 10 lbs (4.5 kg)
|