California Analytical 700 Series Chemiluminescence Analyser
The California Analytical 700 Series CLD Analyzer utilizes the principle of chemiluminescence for analyzing the NO or NOx concentration within a gaseous sample. In the NO mode, the method is based upon the chemiluminescent reaction between ozone and nitric oxide (NO) yielding nitrogen dioxide (NO2) and oxygen. This reaction produces light which has intensity proportional to the mass flow rate of NO2 into the reaction chamber. The light is measured by means of a photodiode and associated amplification electronics. In the NOx mode, NO plus NO2 is determined as above, however, the sample is first routed through the internal NO2 to NO converter which converts the NO2 in the sample to NO. The resultant reaction is directly proportional to the total concentration of NOx.
Features & Benefits
- Measures From Low 3 ppm up to 3,000 ppm Full Scale
- Three Measurement Modes—NO, NOx and NO/NO2/NOx
- Auto Calibration and Ranging
- Fast Response Time
- Electronic Sample and Ozone Flow Control
- Does Not Require Vacuum Pump
- Comprehensive Diagnostics
- Output Options: Voltage, Current, RS-232, TCP/IP, MODBUS
- CE Mark and ETL Listed—Conforms to UL STD 61010-1, Certified to CAN/ CSA C22.2 STD No. 610610.1
- Meets 1065 Requirements
Technical Specifications | |
---|---|
Detector | Chemiluminescence (CLD) Photodiode |
NO/NOx Ranges | Four User-definable from 0-3 to 0-3,000 ppm NO/NOx (Higher Ranges Available upon Request) |
Oxygen | Paramagnetic, 0-25% |
Response Time | Typically < 2 Seconds to 90% Full Scale |
Repeatability | Better than 0.5% of Full Scale |
Linearity | Better than 1% of Full Scale |
Noise | Less than 0.5% of Full Scale |
Zero & Span Drif | Less than 1% of Full Scale per 24 Hours |
CO2 Effect | Less than 2% with 10% CO2 |
Additional Interference Data | CO 1000ppm – N/A HCN 28 ppm – N/A, SO2 500 ppm – N/A NH3 10 ppm – N/A, N2O 201 ppm – N/A |
Flow Control | Electronic Proportional Pressure Controller |
Sample Flow Rate | Typically 1.5 to 2.5 LPM (0.6 LPM with Low Flow Option) |
Converter | Carbon Material @ 205 C; 95 to 100% Efficiency |
Ozonator | Ultraviolet Lamp |
Air or O2 Requirements | Dry Air less than 0.01 ppm NOX at 240 cc/Min. @ 25 psig(Dew Point < -10°C |
NO/NOx Control | Manual/Remote/Auto Cycle (Remote NOx mode by dry contact closure) |